Speaker
Description
Within the polarization program of the NICA collider (Nuclotron-based Ion
Collider fAcility), an absolute polarimeter (APol) is being developed. The setup is
designed for direct measurement of the polarization degree of proton and deuteron
beams in the energy range from 3 to 12 GeV/nucleon. The operating principle of
APol is based on elastic scattering of identical particles (p–p or d–d) on an internal
polarized jet target of hydrogen or deuterium atoms. One of the key parameters of
the source is the beam intensity, which directly determines the density of the atomic
target:
$\rho = I/(\langle v \rangle \cdot s)$,
where 𝐼 is the atomic beam intensity, ⟨𝑣⟩ is the average velocity of atoms in the
beam, and 𝑠 is the beam cross-sectional area. This work considers methods for
producing atomic beams, as well as methods for measuring the profile and intensity
of atomic beams.
To produce an intense atomic beam in the polarized atomic beam source (PABS), a
radio-frequency discharge (13.56 MHz) is used at a pressure of 1–2 mbar in a
borosilicate glass discharge tube with liquid cooling. At optimal absorbed power, the
dissociation degree of H₂/D₂ molecules reaches 90 % due to the dominance of the
dissociative channel at electron energies of 10–20 eV. To suppress atom
recombination on the walls, special surface treatment of the tube and addition of
oxygen (~0.1 %) are applied. Formation of a directed supersonic jet is achieved
through a conical nozzle with temperature stabilization at 60–120 K (accuracy ±0.1
K). During adiabatic expansion, chaotic thermal motion is converted into directed
motion, with a drift velocity of ~1700 m/s for hydrogen.
For diagnostics of the atomic jet parameters, two main methods are used in this work.
The compression tube is designed for absolute measurement of the total atomic beam
intensity, and the quadrupole mass spectrometer (cross-beam QMS) provides
measurement of the dissociation degree and partial fluxes of atomic and molecular
components. The compression tube method is based on passing the atomic beam
through a narrow long tube, the walls of which create diffusion resistance for the
gas. As a result, a stationary pressure proportional to the intensity of the incident
beam is established in the closed volume at the tube outlet, which is measured by a
sensitive pressure gauge (hot-cathode ionization gauge). Calibration of the system
against absolute pressure allows the readings to be converted into the absolute beam
intensity (atoms/s).
The quadrupole mass spectrometer (QMS) determines the dissociation degree and
partial fluxes of atomic and molecular components from ion currents at masses m/𝑒=1 and 2 (for hydrogen) or 2 and 4 (for deuterium), taking into account
ionization cross sections.
The combined use of these methods enables continuous monitoring of the beam
composition and intensity during source tuning. Other diagnostic approaches (wire
calorimeters, visualization on MoO₃ films) are considered promising but are not used
in this work.