Speaker
Description
A new polarimeter has been developed for precision measurements of the polarisation of proton and deuteron beams with total kinetic energies up to 5~MeV/nucleon. The device is designed for the pulsed beam of the linac (30~µs bunch, 5~s period, peak current up to 200~µA) and is based on two well-established polarimeter concepts: the ( ^{3}\mathrm{He}(d,p)^{4}\mathrm{He} ) reaction for simultaneous measurement of vector and tensor polarisation moments (Stephenson \& Haeberli, 1980) and the ( ^{4}\mathrm{He}(d,d)^{4}\mathrm{He} ) elastic scattering for vector polarisation only (Cadmus \& Haeberli, 1975).
The polarimeter uses a gas target cell with a 150~µm Mylar entrance window and 1~µm silicon nitride (SiN) exit windows. The detection system consists of PIN diodes arranged symmetrically around the target, with a solid angle of 3.7~msr. The combination of these two reactions allows for the determination of all four polarisation moments (( t_{11}, t_{20}, t_{21}, t_{22} )) for deuteron beams and the vector polarisation for proton beams.
The design, geometry, and expected performance of the polarimeter are presented. Particular attention is paid to the adaptation of the original continuous-beam designs to the pulsed regime, including the optimisation of the gas pressure and the reduction of the counting rate to safe levels for the detectors. The results from the first beam tests and the calibration procedure are discussed. The polarimeter provides a compact and efficient tool for routine polarisation monitoring and beam tuning at the linac.