Development of control software and a micro-SPECT/CT synchronization subsystem

8 Jun 2026, 14:10
10m

Speaker

Nikita Karpov (men)

Description

This work is devoted to the development of software for controlling a micro-SPECT/CT microtomography system, implemented at the Laboratory of Nuclear Problems (DLNP) of JINR under the scientific supervision of V. A. Rozhkov. The relevance of the study is determined by the need to adhere to the 3R principle (Reduction, Refinement, Replacement), aimed at reducing the number of laboratory animals, as well as the limitations of commercial systems with closed architectures and complex equipment synchronization.
The goal of the work was to create flexible, open-source software based on Python (PyQt5, libximc, Pypixet) that ensures high-precision positioning and synchronization of the Standa mechanical axes with the Timepix semiconductor detector. This study included the development of a modular multithreaded architecture, the implementation of rotational and linear axes calibration algorithms, and scanning protocols with automatic parameter calculation. Particular attention was paid to software synchronization of the event-driven detector with sample movement. Experimental validation confirmed the solution's effectiveness: no positioning errors or mechanical transmission backlash were detected, and high movement repeatability was achieved. A long-term performance test (1.5 hours of continuous scanning) demonstrated the system's stability without interface freezes or communication failures. The developed software saves configurations in JSON format, ensuring the reproducibility of experiments.

Presentation materials