Investigation of the Feasibility of Beam-Based PET Diagnostics in Proton FLASH Therapy

8 Jun 2026, 10:15
10m

Speaker

Darya Mekh

Description

Proton FLASH therapy (dose rate >40 Gy/s) spares healthy tissue but complicates dose monitoring. This work investigates in‑beam PET for range verification in FLASH proton therapy using GEANT4 simulations. A 200 MeV proton beam was directed into a brain‑equivalent phantom, and the production of ¹⁵O, ¹¹C and ¹³N was simulated. The specific activity in the Bragg peak reaches ~915 kBq/mL for clinical FLASH parameters (6×10¹⁰ protons per session), well above the PET detection threshold (~100 kBq/mL). The FLASH pulse duration (0.1 s) is three orders of magnitude shorter than the biological washout time of ¹⁵O, eliminating signal blurring. Under identical proton fluence, the PET signal is identical for FLASH and conventional irradiation, but FLASH provides a distortion‑free snapshot of the dose distribution. In‑beam PET is thus a promising tool for range verification in FLASH proton therapy.

Presentation materials