Study of the Influence of Magnetic Field Orientation and Magnitude on Dose Distributions in Proton Therapy Using Geant4 Simulations

8 Jun 2026, 10:25
10m

Speaker

Daria Prokopenko

Description

The integration of magnetic resonance imaging into proton therapy (MRgPT) offers improved targeting accuracy but introduces dose distortions due to the Lorentz force acting on the proton beam in the MRI magnetic field. This study investigates the influence of magnetic field magnitude and orientation (parallel, perpendicular, and 45°) on the dose distribution of a 200 MeV proton beam in a water phantom using Geant4 Monte Carlo simulations. A voxelized model with 1.0 mm resolution was developed to score three-dimensional dose distributions. Results show that the perpendicular configuration produces the largest lateral Bragg peak displacement and transverse profile asymmetry, while the parallel orientation causes minimal shift but induces beam rotation. The 45° configuration exhibits intermediate effects. All distortions increase with magnetic field strength. The obtained quantitative data can support the validation of dose correction algorithms for MRgPT treatment planning systems.

Presentation materials